44 2033180199

Sixty inappropriate defibrillation shocks in several days due to lead displacement

David Korpas, Jana Haluzikova, Marek Penhaker and Martin Augustynek

Background: Implantable cardioverter-defibrillators (ICDs) are active cardiac implants for immediate treatment of ventricular arrhythmias. They provide life-saving therapy, but may also deliver inappropriate therapy.
Methods and Results : The present report describes a rare occurrence of many inappropriate ICD shocks because of lead displacement shortly after implantation. A 57-year-old overweight man received 353 antitachycardia pacings and 60 shocks of 41 J in total, all within 13 days. The device records show a decrease in the sensed R-waves and pacing impedance fluctuation.
Conclusion: The ventricular electrogram signals recorded and shock channels were typical for lead rubbing.

免責事項: この要約は人工知能ツールを使用して翻訳されており、まだレビューまたは確認されていません。
 
協会、団体、大学向けのピアレビュー出版 pulsus-health-tech
Top